Optical Coherence Tomography and deep Learning for Automated Oocyte Detection in Opaque Follicular Fluid

Chavez-Badiola et al., Optica, 2026


 

Abstract

In in vitro fertilization (IVF), oocytes enclosed in cumulus–oocyte complexes (COCs) must be found in opaque follicular fluid, where transmitted brightfield microscopy (TBM) can miss them. We evaluated whether spectral-domain optical coherence tomography (OCT; 900 nm central wavelength, 3.0 µm axial resolution in air) combined with a real-time detection transformer (RT-DETR) detects oocyte presence within intact bovine COCs.

On an independent test set of 141 wells, with post-denudation confirmation as the reference standard, six embryologists assessing static TBM images detected 65.6–87.0% of oocytes; OCT–RT-DETR detected 99.2% (95% CI 95.8–100%), with 90.0% specificity. OCT-based detection could provide an operator-independent input to automated oocyte search.

Introduction

During in vitro fertilization (IVF), oocytes must be identified in, and isolated from, aspirated follicular fluid, a turbid biological matrix containing blood, fibrin clots, and tissue debris. Oocytes (120 m in diameter) reside within cumulus–oocyte complexes (COCs), cellular structures typically several hundred micrometres in diameter [1] that partially obscure the oocyte within. In a typical IVF setting, oocytes are identified using low-magnification transmitted brightfield microscopy (TBM), which relies on absorption and refractive-index differences to generate contrast. However, TBM provides only surface-level visualization and cannot optically section through opaque or heavily contaminated samples, making detection by the human eye inherently challenging [2]. Recent clinical evidence confirms that this is not a theoretical concern: Mutlu et al. [3] demonstrated that over 54% of IVF patients have oocytes inadvertently discarded during routine manual screening of follicular fluid. Beyond detection, the inability to assess oocyte presence non-invasively within intact COCs represents a fundamental limitation.

Optical coherence tomography (OCT) is a non-invasive imaging modality that generates depth-resolved cross-sectional images using low-coherence interferometry [4–6]. For biological applications, OCT systems typically operate in the near-infrared, e.g. 800 nm – 950 nm, where tissue absorption by water and hemoglobin is low, producing contrast from backscattered light with inherent optical sectioning that enables visualization of subsurface structures inaccessible to TBM.

In reproductive biology, OCT and and its high-resolution variant, optical coherence microscopy (OCM), have been used to image preimplantation embryos [7,8], and predict oocyte developmental potential from intracellular morphology [9,10]. Speckle-variance OCT has been used to detect micron-scale motion as a marker of embryo viability and proposed for assessing damage after cryopreservation [11], and full-field OCT has been used to characterize vitrification-induced changes in mouse embryos [12]. Morawiec et al. [13] integrated a full-field OCT with a commercial fluorescence microscope, enabling visualization of intracellular structures in mouse oocytes and embryos. Trottmann et al. [14] demonstrated that OCT could resolve oocyte structures through intact bovine cumulus layers, with sufficient resolution to evaluate oocyte maturity, but this remained a qualitative observation without automated analysis. To our knowledge, deep learning has not been applied to OCT imaging of oocytes or COCs, despite its proven utility in other areas of reproductive cell analysis [15,16]. Here, we asked whether OCT volumes analyzed by a deep-learning detector classify oocyte presence within intact COCs more accurately than embryologists assessing static TBM images of the same wells. Throughout, detection denotes a per-well binary classification of oocyte presence; physical retrieval of the oocyte was not evaluated. We present a feasibility study comparing embryologist performance on TBM images with automated classification by a Real-Time Detection Transformer (RT-DETR) [17] trained on YZ slices (B-scan-type) extracted from OCT volumetric data of bovine COCs, using post-denudation confirmation as ground truth. Because OCT images through intact cumulus, detection could in future be extended to non-invasive assessment of oocyte maturity, informing the choice between denudation and extended culture. This application was not evaluated here and would require dedicated validation. Furthermore, automation of IVF laboratory procedures is an active and expanding area of development, with each manual step representing both a source of variability and a candidate for instrument-based replacement. A reliable, operator-independent detection step is one prerequisite for automating oocyte search; although instrument-based retrieval is outside the scope of this study.


METHODS

2.1. Study design

This study compared the accuracy of embryologist-performed TBM against an automated system combining OCT with RT-DETR for oocyte detection within bovine COCs, with post-denudation physical confirmation serving as the reference standard.

2.2. Collection of cumulus–oocyte complexes

Bovine ovaries were obtained ex vivo from a local abattoir and transported to the laboratory within 5 hours post-slaughter. Transport was carried out at ambient temperature (20–30 C) without active cooling or heating. Upon arrival, COCs were aspirated from visible follicles (2–8 mm) using a syringe and a 18G needle and immediately placed in a handling medium (HEPES-buffered TCM-199 supplemented with 10% fetal calf serum). Only intact COCs were included in the study. COCs exhibiting diffuse or densely compacted cumulus masses, as well as samples with visible blood contamination, were not excluded in order to reflect challenging real-world laboratory conditions. Selected putative COCs were then individually placed in wells of a BIRR culture dish prefilled with cattle follicular fluid. Putative COCs were identified in the aspirate and transferred to individual wells by an embryologist; the study therefore evaluates detection within isolated putative COCs, not screening of unprocessed aspirates.

COCs were collected in two separate campaigns approximately 12 weeks apart. The development set comprised 69 COCs collected over 2 collection days, and the independent test set 141 wells (131 oocyte-positive, 10 oocyte-negative) collected over 4 collection days. Ovaries were supplied by the abattoir as pooled daily batches without individual animal identification, so the exact number of donor animals cannot be reported; based on routine yields at our facility (approximately 9 ovaries and 36 COCs per collection day), the development and test datasets correspond to approximately 18 and 36 ovaries, respectively. Because ovaries were processed within 5 hours post-slaughter, each donor animal contributed to a single collection day only. No animal, ovary, or dish was represented in both training and test/validation sets, and sibling COCs are therefore necessarily confined to a single dataset.

2.3. Optical coherence tomography and brightfield imaging

The COCs were imaged using a spectral-domain OCT system (Thorlabs Ganymede GAN321 base unit; central wavelength 900 nm) with an LSM02-BB scan lens (effective NA 0.149), giving a lateral resolution at focus of 4.0 m (manufacturer specification; 3.9 m calculated as the spot diameter). Axial resolution was 3.0 m in air (2.2 m in water) and maximum imaging depth 1.9 mm in air (1.4 mm in water); system sensitivity was 89 dB at 80 kHz (106 dB at 1.5 kHz). Volumes were acquired at an A-scan rate of 80 kHz over a lateral field of view of 1 mm 1 mm, sampled at pixels per COC (lateral sampling interval 1.95 m; hereafter a volume), in a measured 5.47 s per volume including galvanometer flyback. The incident optical power at the sample was 1.2 mW; with a spot diameter of 3.9 m, the average power density at focus was W/cm, comparable to that used for optical coherence microscopy of mouse oocytes without reported effects on development [9]. Each volume delivered 6.6 mJ over 1 mm (0.66 J/cm), with a dwell time of 100 s at any point. OCT acquisition was performed immediately after conventional TBM assessment and prior to denudation. Imaging was conducted without the use of contrast agents, fluorescent labels, or fixation, and without physical compression or manipulation of the COCs beyond routine handling.

For imaging, individual COCs were placed in wells of time-lapse culture dishes (BIRR, Culture Dish 35 mm Time Lapse, 12-well format, cat. no. 113031TL) to aid positional stability during image acquisition with two imaging modalities. Each COC occupied a single well of a 12-well dish, such that every volume corresponds to one COC and each dish contained at most 12 COCs; dishes were not reused across collection days, giving a minimum of 6 dishes for the training set and 12 for the test/validation set. TBM and OCT imaging were conducted using a vertically aligned optical configuration, with the COC positioned between the two imaging modalities. A motorized stage was used to precisely control sample positioning and to alternate between TBM and OCT image acquisition without disturbing the COC (Fig.  1 ).


Fig 1.

Schematic representation of the TBM and OCT imaging setup used for COC acquisition.


Volumetric OCT datasets were acquired for each COC, allowing visualization of internal structures within the cumulus cell mass. Acquisition parameters were selected to provide sufficient penetration depth and spatial resolution to identify oocyte-sized structures within dense or blood-contaminated cumulus matrices. All OCT scans were completed following a standardized protocol applied uniformly to all wells regardless of the TBM appearance of individual samples.

Before OCT acquisition, each COC was imaged by TBM, which was also used to locate and centre the sample for OCT. The TBM path comprised a Thorlabs RMS4X-PF objective, an AC508-180-A-ML achromatic tube lens and a Basler acA2500-60uc camera (CMOS, pixels, 4.8 m pixels, global shutter), with illumination from a SmartVision MBL-100x100-WHI backlight and positioning by a Prior ProScan II H30V4 stage (Fig.  1 ). Calibration with a 200 m grid target, cross-checked with a USAF 1951 target, gave a pixel size at the sample of 1.44 m (effective magnification 3.3) and a field of view of mm; the finest element of the target, group 7 element 6 (228 lp/mm; line-pair period 4.4 m), was resolved, so the resolution limit could not be determined beyond this value. This configuration was designed for co-registration with OCT and differs from the stereomicroscopes used for clinical oocyte search, which provide variable zoom and live refocusing.

2.4. Dataset composition and partitioning

A total of 210 OCT volumes were acquired across the study. For every well, a reference standard was established by tracking a complete data triplet: (1) pre-denudation TBM, (2) 3D OCT volumetric scanning, and (3) post-denudation physical confirmation, which served as the ground truth for oocyte presence or absence. Pre-denudation TBM images were preserved only for the test-set wells, as these were required for the independent embryologist assessment described in Section 2.5 ; for training and validation volumes, only the resulting ground-truth label was retained, since model development relied exclusively on the OCT data.

The resulting 210 volumes were partitioned at the volume level into three mutually exclusive sets: training, validation, and test. Volume-level partitioning was used throughout, rather than partitioning at the individual-slice level, to ensure that no information from a given COC was shared across sets (rationale detailed in Section 2.6 ). The training and validation volumes (n = 69) were split following a standard 80–20 ratio, yielding 55 volumes for training and 14 for validation. The remaining 141 volumes were held out exclusively as the test set, used for the final comparative evaluation against human embryologists and never accessed during model development. Table  1 summarizes the composition of each set.

Table 1. Composition and Partitioning of the OCT Volume Dataset

Training, validation, and test volumes underwent distinct slice-extraction procedures, described in Sections 2.6 and 2.7 , respectively.

2.5. Conventional brightfield imaging and embryologist assessment

To ensure a robust and unbiased evaluation, the assessment workflow was structured as follows:

  • Image acquisition: Initial image acquisition and preliminary documentation were performed by a dedicated technician or junior embryologist.

  • Independent labeling: The captured TBM images were then independently reviewed and labeled by two groups of embryologists who were blinded to the OCT findings and the final post-denudation results:

    • Junior embryologists: Individuals in training or with less than 2 years of experience.

    • Senior embryologists: Experienced clinical embryologists with extensive expertise (10+ years) in ART laboratories.

By decoupling image acquisition from the final diagnostic labeling, the risk of confirmation bias during the visual assessment process was minimized.

2.6. Artificial intelligence–based model training

To train a custom RT-DETR object detector, the three-dimensional OCT volumes assigned to training and validation (n = 69; Table  1 ) were processed to extract two-dimensional B-scan YZ slices (Fig.  2 ) of 512 1024 pixels (512 along the Y-axis and 1024 along the Z-axis). Partitioning was performed at the volume level prior to slice extraction, so that all slices originating from a given COC were assigned exclusively to either the training or the validation set. This design choice was necessary to prevent data leakage: slices extracted from the same volume are spatially adjacent and highly correlated, and allowing them to be split across training and validation would have caused the validation set to be contaminated with near-duplicate information already seen during training, artificially inflating performance estimates. Volume-level partitioning ensured that validation slices corresponded exclusively to COCs entirely unseen during training.

Fig 2. Cross-sectional views extracted from an OCT volume. A 3D OCT volume resampled along three orthogonal planes: red (XZ slices), yellow (YZ slices / B-scans), and cyan (en face XY slices). Representative slice stacks are shown at right, color-matched to their cutting planes. YZ B-scans ( px) were used to train the RT-DETR detector. The 3D volume is an illustrative depiction and is not shown to scale; scale bars on the slices, 50 m.

From each training and validation volume, ten YZ slices were extracted within a manually defined region of interest (ROI), uniformly spaced across that range so that the model could learn to identify the oocyte at different lateral positions within the COC. For oocyte-positive volumes, the ROI was bounded by the slice at which the hollow, optically empty cavity corresponding to the space between the cumulus cell mass and the perivitelline space surrounding the oocyte first became visible, and by the slice at which this cavity was no longer discernible. For oocyte-negative volumes, the ROI was bounded by the slice at which the cumulus mass first appeared, and by the slice at which it disappeared. The resulting slices were annotated by embryologists, yielding a YZ-slice dataset of 550 slices for training and 140 slices for validation.

An RT-DETR architecture (Ultralytics) was trained on a computer equipped with an NVIDIA GeForce RTX 4090 graphics processing unit (GPU), using the hyperparameters shown in Table  2 . The detector was RT-DETR-L (Ultralytics). Each slice was normalised to a fixed dB range (17.9-58.7 dB), converted to 8-bit and replicated to three channels; at inference, images were resized internally by Ultralytics, without test-time augmentation. No class weighting, resampling or asymmetric loss was applied; non-annotated regions of each slice served as background.

Table 2. Hyperparameters used to train the RT-DETR detector on YZ B-scan slices

2.7. Artificial intelligence inference and volume-level classification

For the comparative evaluation against human embryologists, the 141-volume test set (Table  1 ) was used. Throughout the inference phase, the artificial intelligence (AI) algorithm remained fully blinded to embryologist classifications and denudation outcomes. For each test volume, all 512 YZ slices were processed by the detector, and detections with a confidence of at least 0.8 were retained. Detections were grouped into tracks across slices: a detection was assigned to an existing track if its bounding-box centre lay within 35 pixels (68 m laterally; Euclidean distance) of that track’s last recorded centroid, and otherwise started a new track. This tolerance accommodates the shift in centroid as the cross-section of a COC changes in size and shape along the slicing axis. Positive slices need not be consecutive: a track was confirmed once it accumulated 20 detections (39 m along the slicing axis), and a volume was classified as oocyte-positive if at least one confirmed track was present. Distances were measured in the original pixels of each slice, with bounding boxes rescaled to native resolution before tracking. All three criteria were fixed on the development data before the test set was evaluated. This procedure yielded a binary classification for each OCT volume (oocyte present or absent), reflecting whether the well should be retained or discarded. These predictions were then cross-referenced with the post-denudation reference standard to determine the model’s diagnostic accuracy. From the start of acquisition to classification, the pipeline took approximately 4.4 min per volume on a workstation with an NVIDIA RTX 4090 GPU, an Intel Core i9-13900K CPU and 64 GB RAM: 5.5 s for acquisition and a mean of 257 s for processing and classification (data extraction 8.4 s; B-scan processing 231.6 s, 0.45 s per B-scan; RT-DETR inference 13.6 s, 0.027 s per B-scan; tracking and I/O 3 s). The pipeline was not optimized for speed.

2.8. Outcome classification and statistical analysis

Final analysis was conducted on the 141-well test set, each well containing a single COC or a sample suspected of containing one. Negative cases, which frequently contained cellular debris or blood, were critical for evaluating each method’s specificity and its ability to distinguish true oocytes from artifacts under challenging conditions. Classification was performed on a per-well basis, aligning with clinical decision-making protocols, where the primary goal is to decide whether a sample should be processed or discarded. For each evaluation method (junior embryologist, senior embryologist, and OCT–RT-DETR), results were categorized against the reference standard established in Section 2.4 as:

  • True Positive (TP): Correct identification of oocyte presence, leading to sample retention.

  • True Negative (TN): Correct identification of an empty well.

  • False Positive (FP): Incorrect identification of an oocyte in an empty well.

  • False Negative (FN): Failure to detect an oocyte, representing the clinical risk of irreversible oocyte loss.

Sensitivity, specificity and accuracy were calculated for each evaluator against the post-denudation reference standard, with exact (Clopper–Pearson) 95% confidence intervals. Cohen’s was computed between each evaluator and the reference standard (not between evaluators), with asymptotic 95% confidence intervals. Because all methods were applied to the same wells, the sensitivity of OCT–RT-DETR was compared with that of each embryologist using an exact McNemar test on the wells where the two methods disagreed. As OCT–RT-DETR missed a single oocyte, these disagreements follow from the counts in Table  3 except for that well, for which we conservatively assumed that each embryologist detected the oocyte. -values were Bonferroni-corrected for six comparisons. Specificity was not compared formally, given the small number of negative wells.

Table 3. Comparative performance metrics for oocyte detection: junior and senior embryologists assessing TBM images versus OCT–RT-DETR (N = 141 wells).

2.9. Ethical considerations

The bovine ovarian tissue used in this study was obtained post-mortem from a local abattoir; therefore, no institutional review board approval for animal research was required. All procedures were conducted in strict accordance with relevant institutional and national guidelines regarding the use of animal-derived biological materials for research purposes and did not involve the use of live animals.

RESULTS

3.1. Study dataset and sample characteristics

A total of 141 individual wells with definitive post-denudation confirmation were included in the final analysis. All samples underwent conventional TBM imaging and three-dimensional OCT imaging prior to denudation.

Post-denudation assessment confirmed the presence of an oocyte in 131 wells (92.9%) and the absence of an oocyte in 10 wells (7.1%). To reflect real-world clinical challenges, the analyzed dataset intentionally included samples with a wide range of morphological characteristics. A substantial proportion of the COCs exhibited diffuse or densely compacted cumulus cell masses. Notably, 69 of 141 wells (48%) showed visible blood contamination or high visual opacity at the time of imaging, conditions that typically hinder oocyte identification under standard microscopy. The OCT–RT-DETR approach agreed with the reference standard more closely ( , 95% CI 0.74–1.00) than any embryologist assessing TBM images ( 0.18–0.40). The individual performance metrics and comparative results for all evaluators are summarized in Table  3 .

Human evaluators across both experience levels missed a clinically significant number of oocytes, with per-observer false negative rates ranging from from 13.0 to 34.4%. The OCT–RT-DETR system reduced this to a single missed oocyte across the entire dataset. Indeed, the sensitivity of OCT–RT-DETR exceeded that of every embryologist (exact McNemar test, Bonferroni-adjusted p 0.001 for all six comparisons).

3.2. Representative Visual Comparison: TBM vs OCT

The clinical challenge of oocyte identification is further illustrated by the representative cases shown in Figs.  3 and 4 . In samples characterized by low cumulus density and the absence of blood contamination (Fig.  3 ), the oocyte is clearly identifiable as a refractive, spherical structure within the cumulus mass. Under these optimal conditions, both human embryologists and the automated OCT–RT-DETR system demonstrated high concordance, as the morphological cues are readily accessible through conventional TBM.

Fig. 3 Representative case of a high-visibility cumulus–oocyte complex (COC). (a) Conventional TBM where the oocyte is clearly identifiable due to low cumulus density and absence of blood contamination. (b) Corresponding 3D-OCT cross-section demonstrating high-contrast resolution of the oocyte’s spherical structure. (c) Physical confirmation of oocyte presence through post-denudation assessment (Ground Truth), (d) Depicts the detection of the AI model.

However, a stark contrast is observed in samples exhibiting high visual opacity due to dense cumulus compaction or significant blood contamination (Fig.  4 ). In these challenging yet common scenarios, the oocyte is virtually "invisible" to the human eye, with the well appearing as a dark, amorphous mass. While such cases frequently led to misclassification and potential oocyte loss by embryologists, the OCT volumetric scan successfully penetrated these opaque layers. By resolving the internal 3D architecture of the COC, the system revealed the oocyte’s high-contrast boundaries and distinct cytoplasmic texture. These cases are consistent with detection failures arising from the optical limits of TBM rather than from operator experience. whereas the OCT–RT-DETR approach provided a reliable mechanism for oocyte rescue in "visually silent" samples.

To further evaluate the spatial correlation between both imaging modalities and the impact of sample opacity on detection, a side-by-side comparison across varying levels of difficulty is presented in Fig.  5 . This multi-column visualization highlights the transition from high-visibility samples to ’visually silent’ cases where the TBM signal is attenuated. In these examples, the oocyte remained resolvable by OCT in wells where TBM images showed no morphological cues.

The single missed oocyte belonged to a volume in which the COC lay unusually high within the imaging depth, well above the dish surface where COCs typically rest. It appeared with low contrast relative to the surrounding speckle, consistent with its distance from the focal plane and with sensitivity roll-off with depth, and produced only 5 confident detections, below the 20 required to confirm a track. COCs in such positions were under-represented in the training data.

Fig. 4. Representative case of oocyte "rescue" in an opaque/blood-contaminated sample. (a) TBM where the sample is optically dense, leading to human misclassification. (b) High-magnification region of interest (ROI) showing no discernible morphological cues under visible light. (c) Corresponding OCT slice of the same ROI, where the volumetric scan successfully penetrates the opaque layers to reveal the hidden oocyte. (d) Physical confirmation after denudation, (e) Depicts the detection of the AI model.

Fig. 5. Spatial correlation and cross-modal superposition across a gradient of visual difficulty. The figure displays representative cases categorized into three levels of identification complexity. (a) Left column- TBM images showing high, moderate, and low oocyte visibility due to increasing cumulus density and blood interference. (b) Middle column- OCT: Corresponding 3D-OCT slices where the oocyte’s spherical structure remains consistently resolved despite the surrounding opacity. (c) Right column- Superposition: overlay of TBM and OCT signals demonstrating the precise anatomical alignment between the two modalities.

DISCUSSION

The results of this study show that automated OCT–RT-DETR detection could support oocyte identification across a range of optical difficulty, although the study evaluated a TBM workflow which does not reproduce clinical search. The comparison therefore quantifies the gap between two image-based workflows, not between OCT–RT-DETR and clinical practice. The contribution of the detector cannot be separated from that of the imaging modality, which would require a blinded reading of the same OCT volumes by embryologists.

In this analysis, human observers failed to identify up to 34.4% of oocytes (false negative rates range 13.0% to 34.4%, six observers), although observers based their assessment on a static TBM image, whilst, in clinical practice, embryologists often improve their chances of detecting an oocyte by diluting contaminated media and physically manipulating putative structures to help identify the presence of "hidden" oocytes. Nonetheless, the reduction to a single false negative with OCT–RT-DETR represents an improvement against 17 for the best-performing observer in this pilot study.

Conventional TBM systems rely on transmitted visible light, which undergoes intense scattering and absorption by dense cumulus cell layers and red blood cells; in optically dense matrices, the oocyte becomes physically inaccessible to human visual assessment. OCT overcomes this limitation by exploiting near-infrared wavelengths and interferometric detection to generate depth-resolved volumetric images with inherent optical sectioning, resolving the oocyte’s spherical structure in the opaque and blood-contaminated wells examined here. Mutlu et al. [3] showed that oocyte loss during routine follicular fluid screening may affect more than half of all IVF patients, indicating that inadvertent discard is common in routine practice. The clinical weight of this finding is greatest in patients with diminished ovarian reserve, where the margin between a successful and a failed cycle may be a single undetected oocyte.

The distribution of human sensitivity across observers warrants specific consideration. Inter-observer variability was not stratified by seniority in a consistent direction: one junior embryologist achieved the highest sensitivity in the cohort, while one senior embryologist recorded the lowest. With six observers, an effect of experience cannot be assessed; however, the pattern observed is consistent with optical inaccessibility limiting detection. The OCT–RT-DETR system performs independently of operator state, fatigue, or experience level, providing a consistent detection floor that could exceed the best human observer.

A defining feature of the oocyte detection problem is the extreme asymmetry of classification error costs. A false negative results in irreversible gamete loss; a false positive carries only a marginal processing time penalty. Sensitivity is therefore the primary optimisation target, with specificity as a secondary constraint. We acknowledge that the absolute number of negative samples in our dataset (10/141) limits the precision of specificity measurement, and prospective studies with larger negative cohorts will be needed to characterise this metric fully. Nevertheless, the OCT–RT-DETR performance profile (99.2% sensitivity, 90.0% specificity) appears skewed in the clinically correct direction, although specificity was assessed on a small sample of n=10 empty wells. This asymmetry should inform future model development: weighted loss functions during training and asymmetric decision thresholds at inference are direct mechanisms for further reducing the false negative rate at an acceptable and bounded cost to specificity.

Beyond potentially improving oocyte detection rates, OCT addresses a more fundamental limitation of current practice. Cumulus cells supply essential nutrients to the oocyte, support maturation through paracrine signalling, and contribute to fertilisation competence [18,19]. Confirming oocyte presence and maturity — indicated by extrusion of the first polar body (15 m) — currently requires denudation, an irreversible act that permanently disrupts the cumulus–oocyte relationship before its biological role is complete, precluding selective management of immature oocytes that might benefit from extended culture under continued cumulus support. Whether OCT can resolve the first polar body through intact cumulus, and so assess maturity before denudation, was not tested here. OCT enables depth-resolved visualisation through intact cumulus layers without physical disruption. Trottmann et al. [14] demonstrated qualitatively that OCT could resolve oocyte structures through intact bovine cumulus; Fluks et al. [9,10] showed that optical coherence microscopy can discriminate oocyte quality and spindle morphology non-invasively in mouse models. The present pipeline required approximately 4.4 min per volume, dominated by B-scan processing rather than inference (Section 2.7 ). It was not optimized for speed; optimizing B-scan processing, and processing B-scans as they are acquired rather than after the full volume is stored, are the main routes to shorter time to classification. Combined detection and maturity assessment before denudation, and allocation of immature oocytes to extended culture, are possible future extensions that will each require dedicated validation.

Several technical limits govern scale-up. First, the study began after an embryologist had isolated each putative COC into its own well, so it evaluates detection, not screening of whole aspirates. A retrieval typically yields 10–90 mL of follicular fluid (1–6 mL per follicle across 10–15 follicles [20]), far more than a 1 mm 1 mm field of view can survey; field of view and volumetric scan time have long limited the use of OCT in embryology. Second, spatial sampling and acquisition speed trade off: enlarging the field of view or increasing pixel density lengthens acquisition proportionally, and faster OCT engines (e.g., 200 kHz) would reduce this at higher system cost. Third, scattering attenuates signal with depth, and COCs lying beyond the imaging depth or floating high in the well fall outside the reliable imaging range; addressing this requires axial repositioning of the scan head, with added time. A practical route may therefore be to locate candidate COCs with a camera and use OCT to confirm the presence of an oocyte in each, rather than to sweep whole dishes.

Several considerations govern clinical translation. Post-imaging viability and development were not assessed: “non-invasive” here denotes the absence of contact, labels and denudation, and the effect of the light exposure reported in Section 2.3 on oocyte competence remains to be tested. This work utilised volume-level thresholds which were fixed a priori; performance across alternative thresholds should be characterized and image contrast and detection performance quantified by degree of blood contamination or cumulus density. Volumetric acquisition requires no contrast agents or sample manipulation beyond routine handling. However, transfer to human samples is the main open question. The bovine COCs studied here were aspirated from unstimulated 2–8 mm antral follicles and mostly carried compact cumulus. Human clinical COCs come from gonadotropin-stimulated pre-ovulatory follicles, typically with expanded, hyaluronan-rich cumulus, and are suspended in larger aspirate volumes with variable blood content and flushing medium. Each difference may alter OCT contrast and scattering, so the present model is validated only for bovine COCs of the morphologies in the test set. We hypothesize that the detector can be adapted to human samples by fine-tuning on a limited human dataset. Testing this requires a prospective study in which human COCs are imaged before denudation within routine handling times, the model is fine-tuned and the decision rule re-fixed on development data only, and performance is measured on a held-out test set with the model frozen and post-denudation confirmation as the reference standard. If validated, operator-independent detection could form one component of automated oocyte search.

Conclusion

OCT combined with a deep-learning detector identified oocytes within intact bovine COCs with higher sensitivity than embryologists assessing static TBM images, including in opaque and blood-contaminated wells where oocytes were most often missed. Because the approach is operator-independent and reproducible, it offers a route to standardized oocyte identification that does not rely on individual expertise. Validation in human samples, a faster processing pipeline and integration with camera-based localization of candidate COCs are the next steps. If these succeed, OCT-based detection could help reduce avoidable oocyte loss in the IVF laboratory, support automated COC search, and open the way to assessing oocytes before denudation in clinical IVF.

 